Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 11: (1 year) CAL VER /5520 INSTRUMENT: Fluke 12: (1 year) CAL VER /5520 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 181 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 DISP Move switch to OFF position. 1.006 DISP 1.006 DISP Note: Erroneous result may occur if switch is not 1.006 DISP placed in OFF position before beginning tests. 1.007 HEAD {RESISTANCE TESTS} 1.008 DISP Move switch to the right position. 1.008 DISP Press SELECT for Ohms. 1.009 HEAD RESISTANCE TESTS: {400 Ohm Range} 1.010 5520 400 0.0Z +0.2U 2W 2.001 5520 400 1.0Z 0.2U 2W 3.001 5520 400 350.0Z 3.4U 2W 4.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 4.002 5520 4 1.000kZ 0.010U 2W 5.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 5.002 5520 40 35.00kZ 0.33U 2W 6.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 6.002 5520 400 100.0kZ 1.0U 2W 7.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 7.002 5520 4 2.900MZ 0.027U 2W 8.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 8.002 5520 40 35.00MZ 0.56U 2W 9.001 HEAD RESISTANCE TESTS: {VCHEK} 9.002 5520 1.960V 0.9% 0.002U 2W 10.001 HEAD {CONTINUITY TEST} 10.002 DISP Press SELECT for Continuity. 10.003 5520 2.000V 0.020U 2W 11.001 HEAD CONTINUITY TEST: {VCHEK} 11.002 5520 3.430V 0.033U 2W 12.001 5520 -0.500V 0.007U 2W 13.001 HEAD {CAPACITANCE TESTS} 13.002 DISP Press uF 13.003 HEAD CAPACITANCE TESTS: {1uF Range} 13.004 5520 1 0.950uF 0.020U 2W 14.001 CON = 14.002 MATH MEM1 = 0.0 14.003 MEMI Enter UUT reading in microfarads: 14.004 MEME 14.005 MEMC 1 uF 0.001U 15.001 HEAD {AC VOLTAGE TESTS} 15.002 DISP Move switch to the center position. 15.002 DISP Press SELECT for VAC. 15.003 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 15.004 DISP Press RANGE to select the 4V range (0.000). 15.005 5520 0.000V S 2W 15.006 MEMI Enter UUT reading in volts AC: 15.007 MEME 15.008 MEMC 4 V +0.003U 16.001 JMPT 18.004 Fluke 11: (1 year) CAL VER /5520 16.002 ASK- Q 16.003 5520 0.000V S 2W 16.004 HEAD AC VOLTAGE TESTS: {4V Range (MAX)} 16.005 DISP Press "M" to select MIN MAX. 16.005 DISP Press "M" again to display MAX. 16.006 MEMI Enter UUT reading in volts: 16.007 MEME 16.008 MEMC 4 V +0.040U 17.001 MEME 17.002 HEAD AC VOLTAGE TESTS: {4V Range (MIN)} 17.003 DISP Press "M" to display MIN. 17.004 MEMI Enter UUT reading in volts: 17.005 MEME # Lower limit should be 0.000 but firmware bug in early units allowed -0.004 17.006 MEMC 4 V -.004U +.040U 18.001 DISP Press "M" for 2s to exit MIN MAX. 18.002 ASK+ Q 18.003 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 18.004 5520 4 0.010V 0.003U 50H SI 2W 19.001 HEAD AC VOLTAGE TESTS: {4V Range (LoZ)} 19.002 DISP Move switch to the right position. 19.003 5520 4 0.010V 0.003U 400H SI 2W 20.001 5520 4 3.500V 0.070U 50H SI 2W 21.001 HEAD AC VOLTAGE TESTS: {4V Range (HiZ)} 21.002 DISP Move switch to the center position. 21.003 5520 4 3.500V 0.070U 400H SI 2W 22.001 HEAD AC VOLTAGE TESTS: {40V Range (HiZ)} 22.002 DISP Press RANGE to select the 40V range (0.00). 22.003 5520 40 35.00V 0.70U 50H SI 2W 23.001 DISP Move switch to the right position. 23.002 HEAD AC VOLTAGE TESTS: {40V Range (LoZ)} 23.003 5520 40 35.00V 0.70U 400H SI 2W 24.001 HEAD AC VOLTAGE TESTS: {400V Range (LoZ)} 24.002 DISP Press RANGE to select the 400V range (0.0). # Warmup 2k resistor so 350V will not current limit 5520A. 24.003 5520 100.0V 80H SI S 2W 24.004 PORT [D1000] 24.005 5520 150.0V 80H SI S 2W 24.006 PORT [D1000] 24.007 5520 200.0V 80H SI S 2W 24.008 PORT [D1000] 24.009 5520 250.0V 80H SI S 2W 24.010 PORT [D1000] 24.011 5520 300.0V 80H SI S 2W 24.012 PORT [D1000] 24.013 5520 400 350.0V 7.0U 80H SI 2W 25.001 HEAD AC VOLTAGE TESTS: {600V Range (HiZ)} 25.002 DISP Move switch to the center position. 25.003 DISP Press RANGE to select the 600V Range (0). 25.004 5520 600 600V 14U 400H SI 2W 26.001 HEAD {DC VOLTAGE TESTS}: 4V Range 26.002 DISP Press SELECT for VDC. 26.002 DISP Press RANGE to select the 4V Range (0.000). 26.003 JMPT 28.003 Fluke 11: (1 year) CAL VER /5520 26.004 ASK- Q 26.005 5520 0.000V S 2W 26.006 HEAD DC VOLTAGE TESTS: {4V Range (MAX)} 26.007 DISP Press "M" to select MIN MAX. 26.007 DISP Press "M" again to display MAX. 26.008 MEMI Enter UUT reading in volts: 26.009 MEME 26.010 MEMC 4 V 0.012U 27.001 MEME 27.002 HEAD DC VOLTAGE TESTS: {4V Range (MIN)} 27.003 DISP Press "M" to display MIN. 27.004 MEMI Use keyboard to enter actual UUT reading in Volts: 27.005 MEME 27.006 MEMC 4 V 0.012U 28.001 DISP Press "M" for 2s to exit MIN MAX. 28.002 ASK+ Q 28.003 HEAD DC VOLTAGE TESTS: {4V Range (LoZ)} 28.004 DISP Move switch to the right position. 28.005 5520 4 0.000V 0.002U 2W 29.001 HEAD DC VOLTAGE TESTS: {40V Range (LoZ)} 29.002 DISP Press RANGE to select the 40V Range (0.00). 29.003 5520 40 0.02V 0.01U 2W 30.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 30.002 DISP Move switch to the center position. 30.003 DISP Press RANGE to select the 400V Range (0.0). 30.004 5520 400 -0.2V 0.1U 2W 31.001 HEAD DC VOLTAGE TESTS: {4V Range (HiZ)} 31.002 DISP Press RANGE for 2s to select autorange. 31.003 5520 4 3.500V 0.034U 2W 32.001 HEAD DC VOLTAGE TESTS: {40V Range (HiZ)} 32.002 5520 40 -35.00V 0.33U 2W 33.001 HEAD DC VOLTAGE TESTS: {400V Range (HiZ)} 33.002 5520 400 100.0V 1.0U 2W 34.001 HEAD DC VOLTAGE TESTS: {600V Range (HiZ)} 34.002 5520 600 -600V 6U 2W 35.001 END